SMITH & NEPHEW, INC. PERI-LOC 4.5MM T25 TI LCK SCR 46MM S-T; PLATE, FIXATION, BONE Back to Search Results
Model Number 71867046
Medical Device Problem Code Fitting Problem (2183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/04/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during surgery, while locking the screw, the screw passed the screw hole on the plate.The operation was finished using the same screw, without surgical delay.The patient was not harmed beyond the described event.
 
Additional Manufacturer Narrative
H3, h6: the device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, a review of this complaint case revealed there were no patient injuries reported.The operation was finished using the same screw.Since no patient injuries are being reported no further clinical/medical assessment is warranted at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible probable cause could include but not limited to procedural variance or user error.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERI-LOC 4.5MM T25 TI LCK SCR 46MM S-T
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11612109
Report Number1020279-2021-02647
Device Sequence Number3496564
Product Code HRS
UDI-Device Identifier03596010641595
UDI-Public03596010641595
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K083032
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number71867046
Device Catalogue Number71867046
Device Lot Number18HT07726
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/08/2021
Supplement Date Received by Manufacturer08/05/2021
Initial Report FDA Received Date04/04/2021
Supplement Report FDA Received Date08/06/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Other; Required Intervention;
-
-