It was reported that during surgery, the back end of the inner sleeve of the humeral 3.5mm hexdriver broke off while the device was outside the patient.The procedure was finished using a backup device from smith and nephew, without surgical delay and no injury to the patient.
|
|
H3, h6: the associated device, used in treatment, was returned and evaluated.A visual inspection was performed and confirmed the humeral hexdriver has a broken tip which was returned with the device.Burrs and gouges indicate signs of excessive wear.A review of the device history records for the listed batch did not reveal any deviation from the standard manufacturing processes.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the complaint history for the listed part revealed no prior complaints for the listed failure mode with the same batch number.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|