• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIOMED GMBH & CO. MEDIZINTECHNIK KG VENOVO VENOUS STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIOMED GMBH & CO. MEDIZINTECHNIK KG VENOVO VENOUS STENT Back to Search Results
Model Number VENUM18100
Medical Device Problem Codes Material Deformation (2976); Activation Failure (3270)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/11/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records will be performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.(expiry date: 10/2022).
 
Event or Problem Description
It was reported that during a stent placement procedure in the left common iliac vein through the femoral vein, the stent delayed in opening and the catheter was stuck on the stent.It was further reported that the stent was deployed upon rotating the handle multiple times but the struts were slightly misaligned.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: the lot history record of this lot was reviewed with special attention to the manufacturing and inspection of this product, and the product was found to have met the specification prior to shipment.Investigation summary: the delivery system was returned for evaluation.The condition of the returned delivery system confirmed an expansion issue.The stent brake of the inner catheter which had been under the stent during deployment was found shifted to distal with heavy imprints, and superficial damage which was considered an indication that the stent adhered to the stent brake after deployment leading to breakaway and shifting upon removal, which leads to a confirmed result for expansion issue.Images demonstrating the placed stent were not available so that the alleged strut misalignment could not be confirmed.Based on the information available the investigation is closed with confirmed result.A definite root cause for the reported event could not be determined.Labeling review: a review of the relevant instructions for use for this product was conducted.The instructions for use was found to address potential complications and adverse events such as incorrect positioning of the stent requiring further stenting or surgery, intimal injury, and malposition.H10: d4 (expiry date: 10/2022), g3.H11: h6 (results) h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during a stent placement procedure in the left common iliac vein through the femoral vein, the stent delayed in opening and the catheter was stuck on the stent.It was further reported that the stent was deployed upon rotating the handle multiple times but the struts were slightly misaligned.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENOVO VENOUS STENT
Common Device Name
VENOUS STENT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
MDR Report Key11612346
Report Number9681442-2021-00201
Device Sequence Number14234217
Product Code QAN
UDI-Device Identifier00801741103728
UDI-Public(01)00801741103728
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P180037
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVENUM18100
Device Catalogue NumberVENUM18100
Device Lot NumberANEX6033
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/01/2021
Initial Date Received by Manufacturer 03/11/2021
Supplement Date Received by Manufacturer04/21/2021
Initial Report FDA Received Date04/05/2021
Supplement Report FDA Received Date05/11/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age48 YR
Patient Weight80
-
-