| Model Number |
1570-02-090 |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Codes |
Fatigue (1849); Foreign Body Reaction (1868); Hypersensitivity/Allergic reaction (1907); Pain (1994); Synovitis (2094); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
01/30/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Initial reporter occupation: lawyer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Unf and medical records received.After review of medical records patient was revised to addressed adverse reaction to metallic debris(armd), pain, systemic symptoms of cognitive decline and pseudotumor.The stem was in 15 degrees anteversion and the trunnion and head showed gross corrosion.The acetabular component was in 45 degrees abduction and anteverted about 20 degrees and had minor lysis.Mri findings reported of significant metal artifact is present for the left hip prosthesis.Doi: (b)(6) 2008 : dor: (b)(6) 2020 (left hip).
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot : a search of the depuy nonconformance (nc) quality system found no nc¿s associated with this product/lot combination.Device history review : a search of the depuy nonconformance (nc) quality system found no nc¿s associated with this product/lot combination.
|
| |
|
Event or Problem Description
|
|
Prior to revision patient alleges stiffness, injury, instability, weakness and metallosis.During revision doctor found out inflamed hypertrophic synovium and anterior capsule was thickened.Doi: (b)(6) 2008 dor: (b)(6) 2020 left hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Appropriate term / code not available (e2402) is used to capture injury (e20), if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|