• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SIG PS CEM FEM W/LUGS LT SZ2.5; SIGMA KNEE PRIMARY : KNEE FEMORAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SIG PS CEM FEM W/LUGS LT SZ2.5; SIGMA KNEE PRIMARY : KNEE FEMORAL Back to Search Results
Model Number 1960-68-500
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Date of Event 03/18/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The patient was revised to address pain.Doi: unknown, dor: (b)(6) 2021, left knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.A worldwide lot specific complaint database search, or device history record (dhr) review, was not possible because the required lot number was not provided.Based on previous investigations, this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the lot code was provided, a manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIG PS CEM FEM W/LUGS LT SZ2.5
Common Device Name
SIGMA KNEE PRIMARY : KNEE FEMORAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key11613648
Report Number1818910-2021-06964
Device Sequence Number3160354
Product Code JWH
UDI-Device Identifier10603295075349
UDI-Public10603295075349
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K073529
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1960-68-500
Device Catalogue Number196068500
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/18/2021
Supplement Date Received by Manufacturer04/29/2021
Initial Report FDA Received Date04/05/2021
Supplement Report FDA Received Date05/04/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SIG MOD TIB TRAY CEM COCR 2.5; SIG PS CEM FEM W/LUGS LT SZ2.5; SIGMA STAB XLK INS 2.5 12.5MM; SIG MOD TIB TRAY CEM COCR 2.5; SIG PS CEM FEM W/LUGS LT SZ2.5; SIGMA STAB XLK INS 2.5 12.5MM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
-
-