| Model Number |
7213 |
| Medical Device Problem Code |
Device Markings/Labelling Problem (2911)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/09/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device returned to manufacturer: the packaging, shaft, hypotube, tip and balloon were microscopically and visually examined.The device was received in a carton for batch 26757951.There was no pouch in the carton.The device was from batch 26757951.The device inspection found there were numerous kinks.There was contrast in the inflation lumen and balloon.The balloon was loosely folded.Microscopic inspection revealed tip damage.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis did not confirm the reported mislabeling as there was no pouch returned with the device.
|
| |
|
Event or Problem Description
|
|
It was reported that a device had an incorrect label.A physician had requested a 3.5 x 15mm nc emerge balloon catheter.The sealed box was labelled as 3.5 x 15mm.The sealed inner pouch was labelled 4 x 8mm.The actual balloon was a 3.5 x 15mm despite the labelling on the inner pouch.The device was used for the procedure, which was completed successfully with no patient injury.
|
| |
|