MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION TUR IRRIGATION SET; SYSTEM, IRRIGATION, UROLOGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION TUR IRRIGATION SET; SYSTEM, IRRIGATION, UROLOGICAL Back to Search Results
Catalog Number JC4041P
Device Problem Incomplete or Inadequate Connection (4037)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Device manufacturer address: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a customer was having difficulty spiking a y-type tur/bladder irrigation set into a bag as it was alleged that the spike was too short and less sharp than usual.The event was further described as the spike not spiking far enough into a 3000cc baxter bag during insertion.This was observed during an unspecified process step.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Additional information was added to d9, h3 and h6.H10: two (2) actual samples were received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional testing was performed including pressure testing and clear passage testing, and the results were satisfactory.Additionally, priming testing was performed, and no defect was observed.Dimensional testing was also performed, and the spike was according to specification.The reported condition was not verified in both samples.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TUR IRRIGATION SET
Type of Device
SYSTEM, IRRIGATION, UROLOGICAL
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11613942
MDR Text Key246602457
Report Number1416980-2021-01856
Device Sequence Number1
Product Code LJH
Combination Product (y/n)Y
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 05/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/05/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberJC4041P
Device Lot NumberASKU
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/30/2021
Date Manufacturer Received05/05/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
UNSPECIFIED BAXTER BAG
-
-