| Model Number |
N/A |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/22/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the bearings were discovered missing from the tasp during the cleaning process.There was no patient involvement.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and corrected information.The complaint sample was evaluated and the reported event was confirmed through physical evaluation.The returned shim exhibited signs of repeated use (nicked/gouged) and both of each of the springs and ball bearings was disassembled and not returned for evaluation.The device history records were reviewed and no discrepancies were identified.A corrective and preventative action investigation into this issue was initiated to recommend against using ultrasonic cleaning as a sterilizing technique for these devices.The device was subsequently re-designed to account for this failure mode with a new locking mechanism.This device was manufactured prior to the design modification and therefore the root cause is the previously addressed design issue.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|