SYNTHES GMBH CHUCK W/KEY F/COMPACT AIR DRIVE+POWER DR; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 511.730 |
Device Problem
Compatibility Problem (2960)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional narrative: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the device damages/breaks the k-wire device.Therefore, the reported condition was confirmed.The assignable root cause of this condition was determined to be traced to maintenance, which is user error/misuse/abuse.Udi (b)(4).
|
|
Event Description
|
It was reported from (b)(6) that during service and evaluation, it was determined that the chuck with key device damages/breaks the k-wire device, the coupling could not engage the attachment device, and there was component damage.It was further determined that the device failed pretest for check handpiece coupling, and check clamping range of the drill chuck.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported, however, it was reported that the event occurred in 2021.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|