Catalog Number 511.730
Device Problem Compatibility Problem (2960)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2021
Event Type  malfunction  
Manufacturer Narrative
If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional narrative: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the device damages/breaks the k-wire device.Therefore, the reported condition was confirmed.The assignable root cause of this condition was determined to be traced to maintenance, which is user error/misuse/abuse.Udi (b)(4).
 
Event Description
It was reported from (b)(6) that during service and evaluation, it was determined that the chuck with key device damages/breaks the k-wire device, the coupling could not engage the attachment device, and there was component damage.It was further determined that the device failed pretest for check handpiece coupling, and check clamping range of the drill chuck.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported, however, it was reported that the event occurred in 2021.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHUCK W/KEY F/COMPACT AIR DRIVE+POWER DR
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer Contact
kara ditty-bovard
eimattstrasse 3
oberdorf 4436
SZ   4436
6107428552
MDR Report Key11614638
MDR Text Key243938194
Report Number8030965-2021-02472
Device Sequence Number1
Product Code HWE
UDI-Device Identifier07611819036420
UDI-Public7611819036420
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 03/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/05/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number511.730
Device Lot Number1037919
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/19/2021
Date Manufacturer Received03/23/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/04/1999
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-