|
There is more information on the device evaluation.This supplemental report is being submitted to provide this information.The device history record review confirmed that device has no abnormalities, special adoption, or variations in manufacturing.Though there can be a possibility of peeling off of the forceps cover glue due to aging, but considering that there is general damage concentrated on the distal end, the cause is likely due to external force such as strong rubbing on the area in normal use including reprocess.The instructions for use includes the following statements: inspection of the endoscope: 3.Inspect the external surface of the entire insertion section including the bending section and the distal end for dents, bulges, swelling, scratches, holes, sagging, transformation, bends, adhesion of foreign bodies, missing parts, protruding objects, or other irregularities.
|