ALLERGAN (COSTA RICA) STYLE 410 COHESIVE SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Model Number |
MX-410325 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Inflammation (1932)
|
| Date of Event |
07/24/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
In response to fda report number: mw5091025.A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: visual analysis of the returned device identified the weight to the specification, a crease fold and deformation.Cloudiness and bubbles were observed after the autoclave disinfection process.Based on the device analysis the final assessment is: no issues found related with the manufacturing process.The event of irritation/inflammation is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: irritation/inflammation.Allergan did not submit this mdr within 30 days of becoming aware.Recent stimulated reporting related to 2011068-7/2/19-001-r has increased complaint and mdr volume.Allergan is implementing a plan to address the increased volumes.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a right side implant exchange of textured device due to "increased risk for alcl." patient also reported "pain and burning in upper body, pain more consistently in my right breast" and "chronic inflammation." patient additionally reported "discomfort and muscle and joint paint as well" and "autoimmune system has been compromised" which are not device related.Healthcare professional additionally reported capsular contracture, baker grade ii.Device has been explanted.
|
| |
|
Search Alerts/Recalls
|
|
|