• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US LIGHT SLOTTED HAMMER; KNEE INSTRUMENT : IMPACTION INSTRUMENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US LIGHT SLOTTED HAMMER; KNEE INSTRUMENT : IMPACTION INSTRUMENTS Back to Search Results
Model Number 2421-22-000
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/30/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2021, the patient underwent the artificial bone head surgery with the slotted hammer in question.During the surgery, the white nylon part of the hammer came off.The surgeon tried to attach the nylon part to the hammer, but he couldn¿t.The surgery was completed successfully without any surgical delay.No further information is available.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative:  added: d9.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: h3.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : examination of the returned device does not confirm device breakage, but did find the the device is stripped and cannot be assembled.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGHT SLOTTED HAMMER
Common Device Name
KNEE INSTRUMENT : IMPACTION INSTRUMENTS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key11615340
Report Number1818910-2021-07040
Device Sequence Number3052441
Product Code FZY
UDI-Device Identifier10603295122906
UDI-Public10603295122906
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 03/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2421-22-000
Device Catalogue Number242122000
Device Lot NumberJG715A35
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/27/2021
Initial Date Received by Manufacturer 03/29/2021
Supplement Date Received by Manufacturer03/29/2021
04/27/2021
06/07/2021
Initial Report FDA Received Date04/05/2021
Supplement Report FDA Received Date04/09/2021
05/04/2021
06/07/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-