SYNTHES GMBH ELECTRIC PEN DRIVE 60,000 RPM; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 05.001.010 |
Device Problems
Unintended System Motion (1430); Fail-Safe Did Not Operate (4046)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the reported condition was confirmed.The assignable root cause was determined to be traced to component failure from normal wear.Udi: (b)(4).
|
|
Event Description
|
It was reported from (b)(6) that during service and repair/pre-testing, it was observed that the electric pend drive device had low power, ran in the locked position, and had unintended motion.It was further determined that the device failed pretest for check speed with tachometer, check power with power test bench, functional test fwd (forward) and rev (reverse) running, check for unintended motion, and check function with test hand switch.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The date of event was unknown but was known to have occurred in 2021.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|