• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION MAXFORCE TTS; DILATOR, ESOPHAGEAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION MAXFORCE TTS; DILATOR, ESOPHAGEAL Back to Search Results
Model Number M00589180
Medical Device Problem Codes Material Puncture/Hole (1504); Defective Device (2588)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: date of event was approximated to (b)(6) 2021 as no event date was reported.(b)(6).(b)(4).Investigation result: a visual examination of the returned complaint device found that the balloon did not have any visual defects and was in a good condition.The catheter of the device was carefully inspected and no damages were found.Functional analysis was performed, and the balloon was inflated without a problem; however, the balloon would not hold pressure due to a pinhole in the proximal section on the body of the balloon.It is possible that the factors encountered during the procedure such as the technique used by the physician and/or the interaction with the scope could have caused the balloon pinhole.Therefore, the most probable root cause is adverse event related to procedure.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.A labeling review was performed and from the information available, this device was used per the directions for use (dfu)/product label.
 
Event or Problem Description
It was reported to boston scientific corporation that a maxforce tts dilatation balloon was used in a procedure performed on an unknown date.It was reported that the device was defective.No patient complications were reported as a result of this event.Boston scientific has been unable to obtain additional information regarding the event to date, despite good faith efforts.Investigation results revealed that the balloon had a pinhole; therefore, this is now an mdr reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAXFORCE TTS
Common Device Name
DILATOR, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key11616191
Report Number3005099803-2021-01458
Device Sequence Number2998691
Product Code KNQ
UDI-Device Identifier08714729099925
UDI-Public08714729099925
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K934697
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/16/2022
Device Model NumberM00589180
Device Catalogue Number8918
Device Lot Number0025219935
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/05/2021
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/12/2021
Initial Report FDA Received Date04/05/2021
Patient Sequence Number1
-
-