| Model Number |
IPN045176 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/19/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that "battery compartment cover opened and battery fell out during use".No report of patient injury or harm.Patient condition unknown at time of report.
|
| |
|
Event or Problem Description
|
|
It was reported that "battery compartment cover opened and battery fell out".No report of patient injury or harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).Additional information received indicates the event occurred when assembling the blade to the handle, prior to use on a patient.It was also reported that the patient condition is "ok".The actual sample was not returned; however, the customer provided two photos for evaluation.The manufacturing site reports "from the defect pictures, we could find that the spring is deformed and this indicated that the cap is rotated from disassembly or by mistake.Assembly part's spring is straight.The handle product is a disposable device, for one time use only.If user/someone rotate the battery cap by mistake or on purpose (for example: take the battery), then the device could not be assembled and used any more." without the actual device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
|
| |
|