DEPUY SYNTHES PRODUCTS LLC BEARING SLEEVE, 15CM CURVED; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
|
Back to Search Results |
|
| Model Number |
MA-15C |
| Medical Device Problem Code |
Excessive Heating (4030)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.However, since the investigation is still on-going, the assignable root cause could not be determined at this time.Once the investigation has been completed, a supplemental medwatch will be submitted accordingly.Udi: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during service and repair pre-testing, it was discovered that the bearing sleeve device was overheating.The event was not related to surgery.There was no patient involvement.There were no injuries or medical intervention associated with this event.The exact date of the event was not reported.However, it was reported that the event occurred in 2021.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: device evaluation: this device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Quality engineering evaluated the device and it was determined that the bearing sleeve device exhibited heat.It was further determined that the device failed pretest for max temperature of sleeve.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to component failure from normal wear.
|
| |
|