Catalog Number 7N8378
Device Problems Packaging Problem (3007); Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/09/2021
Event Type  malfunction  
Manufacturer Narrative
Initial reporter phone no.(b)(6).Device manufacturer address 1: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that two (2) one-link non-dehp standard bore catheter extension sets had sterile packaging issues.During setup/preparation, it was discovered the two sets were received conjoined.This issue was further described as half of one set was protruding outside of its sterile packaging and was within the packaging of the second set.The tubing of the set that was between the two packages was also cut.This event was identified during setup/preparation prior to patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information was added to d9, h3, h4 and h6.H10: two (2) actual samples were received for evaluation.Visual inspection was performed using the naked eye which observed that one set passed through the package into another package.The exposed tubing was cut.The cause of the cut was from the blades on the packaging machine.The reported conditions were verified.The cause of the conditions were due to improper packaging operation during manufacturing.A nonconformance has been opened to address the product placement issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTENSION SETS WITH ONE-LINK NEEDLE-FREE LV CONNECTOR
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11617016
MDR Text Key243928735
Report Number1416980-2021-01888
Device Sequence Number1
Product Code FPA
UDI-Device Identifier00085412091792
UDI-Public(01)00085412091792
Combination Product (y/n)Y
PMA/PMN Number
K132734
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 05/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number7N8378
Device Lot NumberR19L04163
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/13/2021
Initial Date Manufacturer Received 03/10/2021
Initial Date FDA Received04/05/2021
Supplement Dates Manufacturer Received05/07/2021
Supplement Dates FDA Received05/11/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-