SMITH & NEPHEW ORTHOPAEDICS AG EXTRAMEDULLARY ALIGNMENT ROD; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
| Model Number |
114861 |
| Medical Device Problem Codes |
Corroded (1131); Material Discolored (1170)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/16/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Complaint reference: (b)(4).The device, used in treatment, was returned for evaluation.A visual inspection of the returned device confirmed the stated failure mode.The device was found to be rusty and discolored, rendering the device inoperable., rendering the device inoperable.The device shows signs of extensive use.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary. at this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Event or Problem Description
|
|
It was reported that during a field inspection the ipg mis dcf ap ct blk 3 was noted to be rusty and discolored, the extramedular alignment rod was broken, the gii quick connect handle dented and did not retract, the ipg ps high flex impctr hd 1-2 gii ps high flex impctr hd 3-8, gii tibial base impactor, gii narrow pcl retractor, 11mm gns ii tibial drill with burrs and gouges and outer housing cap with one component missing and broke.No case reported; therefore, there was no patient participation.
|
| |
|
Search Alerts/Recalls
|
|
|