| Catalog Number |
EXACTA24DY |
| Medical Device Problem Code |
Microbial Contamination of Device (2303)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/10/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial reporter address: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the display of an exactamix automated compounding device was contaminated with a microbial growth.The contamination was observed during an unspecified process step prior to patient use.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, h3, h6.Correction to g1 address.H10: the device was received for evaluation.A visual inspection was performed which noted the device had unknown fluids on the display.Functional system level testing was performed with no issues noted.The reported microbial growth was not verified; however, fluids were found on the display.The fluid on the display is outside the fluid path and would not impact the sterility of the solution compounded.The cause of the fluids could not be determined.The display would be serviced and replaced.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|