| Model Number |
4050 |
| Medical Device Problem Codes |
Incorrect Measurement (1383); Sensing Intermittently (1558)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/18/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
|
| |
|
Event or Problem Description
|
|
The customer reported the sensor is intermittently seeing baseline wander and intermittently working.No patient impact or consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional manufacuring narrative: the returned sensor was evaluated.During evaluation the sensor passed all visual and functional testing. during simulation testing, the sensor passed manual and preset conditions and provided accurate measurements.The sensor was determined to be functioning as designed. .
|
| |
|
Event or Problem Description
|
|
The customer reported the sensor is intermittently seeing baseline wander and intermittently working.No patient impact or consequences were reported.
|
| |
|
Search Alerts/Recalls
|