| Model Number |
TRULIANT PIN PULLER SYRINGE STYLE |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/17/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Event or Problem Description
|
|
As reported, the truliant tibia pin puller had a prong break-off when a headed pin was trying to be removed from the bone within the trial tibia tray.The surgeon was able to get the pin out of the bone using a backup pin puller that was peel packed.There was no delay in the surgery.The patient left the or stable.There were no pieces left in the patient.The broken pin puller and the broken prong were located, identified, and isolated.Patient was last known to be in stable condition following the event.The device is available for return to the company.
|
| |
|
Additional Manufacturer Narrative
|
|
(h3) the evaluation noted that according to the manufacturer investigation, the device history records of the subject devices were reviewed, and no issues were observed.A design review showed that the breakage area of the component is relatively thin which does not offer great safety margin when the instrument is subject to off-axis loading.Although the evaluation couldn¿t confirm user misuse, off-axis loading which can be generated by the user applying a bending moment to the device during pin driving/pulling could contribute to the breakage.The most likely root cause associated with the reported event are the breakage area of the component being relatively thin which does not offer great safety margin when the instrument is subject to off-axis loading and/or off-axis loading when the user applies a bending moment to the device during pin driving/pulling.As a correct action, manufacturer have update design to improve strength of breakage area and update operative technique to instruct users to avoid using the instrument off-axis.A risk assessment has been initiated to escalate this issue.
|
| |
|
Search Alerts/Recalls
|