• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SCREW BIOSURE REGENESORB 10MM X 25MM; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SCREW BIOSURE REGENESORB 10MM X 25MM; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE Back to Search Results
Model Number 72204408
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/17/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that, during a procedure, the biosure screw broke while fixing.No delay and a back up was available to complete the procedure.No other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
H10 h3, h6: the reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was an isolated event.A review of risk management files found that the reported failure was documented appropriately.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of the polymer found that the storage requirements, material specifications, and material tests were appropriately documented.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
H10: h3, h6: the reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A review of the device records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was an isolated event.A review of risk management files found that the reported failure was documented appropriately.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of the polymer found that the storage requirements, material specifications, and material tests were appropriately documented.The device was returned in the original packaging.The returned screw was fractured on the distal tip.Based on the condition of the product material found during visual inspection, additional material testing is not required.The complaint was confirmed.Factors that could have contributed to the reported event include excessive force on the device, excessive torque on the device, attempted correction of a damaged device, off-axis insertion, improper preparation of the insertion site, or an inadvertent impact event.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCREW BIOSURE REGENESORB 10MM X 25MM
Common Device Name
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
MDR Report Key11625585
Report Number1219602-2021-00787
Device Sequence Number3356737
Product Code MAI
UDI-Device Identifier00885554036848
UDI-Public00885554036848
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K142948
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/07/2022
Device Model Number72204408
Device Catalogue Number72204408
Device Lot Number50838438
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/02/2021
Initial Date Received by Manufacturer 03/18/2021
Supplement Date Received by Manufacturer05/05/2021
11/02/2021
Initial Report FDA Received Date04/06/2021
Supplement Report FDA Received Date05/06/2021
11/03/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/07/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
-
-