| Model Number |
7770823 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/15/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Information was received from a healthcare facility via a medtronic representative regarding a patient implanted with a spacer for an unknown spinal indication.It was reported that the peek portion of the implant made contact with the superior portion of the endplate, causing it to snap off the metal portion of the elevate cage.The spacer broke upon insertion.There were no patient symptoms or complications as a result of the event.The date of implant and explant was (b)(6) 2021.The product was replaced by a medtronic product.No further complications were reported/ anticipated.
|
| |
|
Event or Problem Description
|
|
Information was received from a healthcare facility via a medtronic representative regarding a patient implanted with a spacer for a tlif.It was reported that the peek portion of the implant made contact with the superior portion of the endplate, causing it to snap off the metal portion of the elevate cage.The spacer broke upon insertion.There were no patient symptoms or complications as a result of the event.The date of implant and explant was (b)(6) 2021.The pre-op diagnosis and levels implanted are unknown.There was no delay in surgery time or patient harm.The product was replaced by a medtronic product.No further complications were reported/ anticipated.
|
| |
|
Additional Manufacturer Narrative
|
|
B5: updated event details h3: product analysis part# 7770823, lot# 0792724w- visual and optical inspection confirmed the plastic portion of the spacer has broken off.It appears the hinges have been overload.This type of damage is consistent with material overload.H6: updated eval.Code method and eval.Code result post analysis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|