| Model Number |
DCB00 |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/22/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If explanted, give date: not applicable, as the lens remains implanted. device evaluation: product testing could not be performed since the product was not returned for evaluation.Therefore, the reported issue could not be verified, and product quality deficiency could not be determined.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.A search revealed that no additional complaint was received from this production order.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the cartridge tip of the preloaded device was cracked during the insertion of the lens into the eye.The lens was fully implanted and the event was observed after the implantation.The outcome does not significantly interferes with activities of daily life.No further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d9: device available for evaluation: yes returned to manufacturer: may 11, 2021 section d9: returned to manufacturer on: yes device evaluation: the product was returned to the manufacturing site for evaluation.The plunger and pushrod was observed in advanced position residues of viscoelastic material was observed on cartridge.The cartridge tip was observed deformed and partially crack.The lens was not returned, it was implanted.Based on the sample evaluation, there is no evidence to suggest that the complaint unit has been affected by the manufacturing process.The condition in which the sample returned is consistent with a product that was handled and prepare for surgical process.The complaint issue reported was verified.Based on the analysis of the returned product, there is no indication of product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|