| Model Number |
RONYX25038UX |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/31/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
One resolute onyx coronary drug eluting stent was attempted to be used during a procedure to treat a mildly tortuous, moderately calcified lesion with 90% stenosis in the mid lad.The device was inspected prior to use with no issues noted.Negative prep was not performed.The lesion was pre-dilated.The device did not pass through a previously deployed stent.Resistance was encountered when advancing the device but excessive force was not used during delivery.It was reported that stent deformation occurred in vivo during positioning.The procedure was completed using another resolute onyx stent.No patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: device returned for evaluation.The stent was positioned on the balloon between the marker bands as per specifications.Deformation was evident to the proximal stent wraps with struts raised.Although the stent meets od dimensional measurement criteria, stent deformation is confirmed based on the failed visual inspection.No deformation was evident to the distal tip.The inner lumen patency was verified with a 0.015 inch mandrel.No other damage evident to the remainder of the device.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|