MAUDE Adverse Event Report: DATEX-OHMEDA, INC. CARESCAPE R860; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATEX-OHMEDA, INC. CARESCAPE R860; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Device Problem Therapeutic or Diagnostic Output Failure (3023)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/09/2021
Event Type  malfunction  
Manufacturer Narrative
Patient information could not be obtained due to country privacy laws.The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.Unique identifier: (b)(4).A ge healthcare service representative performed a checkout of the system and confirmed the reported issue.The patient valve and safety guard filter were replaced to resolve the issue.
 
Event Description
The hospital reported the unit had an error that results in a loss of mechanical ventilation.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARESCAPE R860
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer Contact
john szalinski
3030 ohmeda drive
madison, WI 
MDR Report Key11626775
MDR Text Key244201767
Report Number2112667-2021-00956
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
PMA/PMN Number
K172045
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial
Report Date 04/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/07/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/09/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/02/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-