H10: the fda rn number for the mdr 2020394-2021-00817 was inadvertently submitted as 2020394.The correct fda rn number was 1018233.H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: five needle was returned for evaluation, this complaint will account for one unit.The components received included cannulas, stylets, with needle protectors; the stylets were returned inserted within the cannulas.The packaging appeared to be open at the proximal end.The device components appeared unaffected.Based on the findings, the investigation is confirmed for the reported packaging issue as the it was noted that the packaging was opened in conditions.A definitive root cause for the reported packaging issue could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 02/2023),.
|