| Model Number |
1960-50-450 |
| Medical Device Problem Code |
Device Contaminated During Manufacture or Shipping (2969)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/17/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Intraoperatively the implant was found to have adhesions of sterile packaging material on the articulating surface.Another implant was available and used without issues.No surgical delay, no adverse patient consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # : (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: examination of the returned device found a foreign material on the device.Root cause and corrective action are documented in the capa system.This device was manufactured on 24-feb-2020.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: 1) quantity manufactured: (b)(4).Date of manufacture: 2/24/2020.Any anomalies or deviations identified in dhr: none.Expiry date: 1/31/2030.5) ifu reference: 090200252 rev m.
|
| |
|
Search Alerts/Recalls
|