| Model Number |
RBY4C0835-B |
| Medical Device Problem Code |
Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/10/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a coil embolization procedure in the ovarian vein using a ruby coil and a lantern delivery microcatheter (lantern).During the procedure, after advancing approximately five centimeters of a ruby coil into the lantern, the ruby coil became stuck.Therefore, the ruby coil was removed.The procedure was completed using a new ruby coil and the same lantern.There was no report of an adverse effect to the patient.
|
| |
|