MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number 3545107MC
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Pain (1994)
Date of Event 03/02/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: left breast implant rupture, and discomfort.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a (b)(6) year-old caucasian female patient underwent revision breast reconstruction surgery with a 510cc mentor memorygel breast implant and experienced breast implant rupture, and discomfort on her left side postoperatively.As a result, the patient underwent left side explantation and replacement with catalog number 3505001bc; serial number (b)(4) on (b)(6) 2021.
 
Additional Manufacturer Narrative
On (b)(6) 2021, the mentor failure analysis lab received the device for evaluation.The analysis has begun, but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On april 26, 2021, device evaluation was completed.Device evaluation summary: mentor conducted a visual inspection and leak testing of the returned device.Visual analysis of the returned sample revealed that no damage or anomalies were observed on the ssx mod classic gel, 510cc breast implant.Leak testing was performed, in accordance with mentor procedures, and no leak sites were detected during the analysis.Although no conclusion could be reached on the cause of the reported event, the instructions for use contain the following caution: rupture status identified by mri evaluation includes both suspected ruptures, which are those ruptures identified by mri but not confirmed by explantation and examination of the device, and confirmed ruptures, which are those ruptures that are confirmed by evaluation of the explanted devices.The event described could not be confirmed as the breast implant was returned without detectable damage.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Most women undergoing augmentation or reconstruction with a mammary prosthesis will experience some pain postoperatively.While pain normally subsides in most women as they heal from surgery, it can become a chronic problem in other women.Chronic pain can be associated with a variety of factors.Surgeons should instruct their patients to inform them if there is significant pain or if pain persists.Pain is a known complication associated with these devices and is referenced in our current product insert data sheet.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
MDR Report Key11628506
Report Number1645337-2021-03750
Device Sequence Number11583290
Product Code FTR
UDI-Device Identifier00081317004493
UDI-Public00081317004493
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2019
Device Model Number3545107MC
Device Catalogue Number3545107MC
Device Lot Number6816892
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/13/2021
Initial Date Received by Manufacturer 03/24/2021
Supplement Date Received by Manufacturer04/13/2021
04/26/2021
Initial Report FDA Received Date04/07/2021
Supplement Report FDA Received Date04/14/2021
04/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age43 YR
-
-