• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER K2 EDTA (K2E) 18 MG BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER K2 EDTA (K2E) 18 MG BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Model Number 368589
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/16/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported the bd vacutainer® k2 edta (k2e) 18 mg blood collection tubes experienced discolored / abnormal additive form.The following information was provided by the initial reporter.The customer stated: it was reported that the plasma looks like jelly."i am reaching out as related to the 10 ml k2edta tube, lot #0167257, exp 06-30-22.We are seeing issues now with approximately 6-7 specimens in which the plasma is described as being like jelly.Have you heard of any issues of this nature from anywhere else.We are wondering if it is this specific lot number alone, as this is the only lot number we currently have on hand for both of our hospital campuses.".
 
Event or Problem Description
It was reported the bd vacutainer® k2 edta (k2e) 18 mg blood collection tubes experienced discolored / abnormal additive form.The following information was provided by the initial reporter.The customer stated: it was reported that the plasma looks like jelly."i am reaching out as related to the 10 ml k2edta tube, lot #0167257, exp 06-30-22.We are seeing issues now with approximately 6-7 specimens in which the plasma is described as being like jelly.Have you heard of any issues of this nature from anywhere else.We are wondering if it is this specific lot number alone, as this is the only lot number we currently have on hand for both of our hospital campuses.".
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d10: device available for eval yes, d10: returned to manufacturer on: 2021-03-29.H6: investigation summary : bd received 10 contaminated returned samples.Testing cannot be performed on used samples.Therefore, retention samples from bd inventory were evaluated for any additive issues.All testing was within specification.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.Bd was not able to identify a root cause for the indicated failure mode.H3 other text : see h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER K2 EDTA (K2E) 18 MG BLOOD COLLECTION TUBES
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
MDR Report Key11628541
Report Number1917413-2021-00270
Device Sequence Number19749807
Product Code JKA
UDI-Device Identifier50382903685894
UDI-Public50382903685894
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
BK980011
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date06/30/2022
Device Model Number368589
Device Catalogue Number368589
Device Lot Number0167257
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/29/2021
Initial Date Received by Manufacturer 03/17/2021
Supplement Date Received by Manufacturer04/13/2021
Initial Report FDA Received Date04/07/2021
Supplement Report FDA Received Date04/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-