| Model Number |
TV-IL1422140-J |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
03/16/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device involved in the event will not be returned for evaluation as it remains implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by the clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
The patient was initially implanted with the ovation ix abdominal stent graft system to treat an abdominal aortic aneurysm (aaa).Approximately (4) four years post initial implant the patient was identified with a type 1b endoleak on the right side.The physician elected to implant three (3) iliac limb stent grafts to resolve the reported event.The patient is doing well post procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.The review confirms there were no manufacturing or processing non-conformities identified that would contribute to the reported adverse event/incident.There are no other equivalent adverse events/incidents for this lot number existing within the endologix complaint handling system.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows the reported type ib endoleak is unconfirmed due to a lack of the medical information.This is not consistent with the reported adverse event/incident.Procedure-related harms, device, user, procedure, or anatomy relatedness of this event could not be determined with the medical records available for review.The final patient status was reported as doing well post secondary endovascular procedure.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Corrections: h6 evaluation result codes; remove 3233.H6 evaluation conclusion codes; remove 11.
|
| |
|
Search Alerts/Recalls
|