| Model Number |
123522A |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
03/15/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the balloon of the foley catheter burst inside the patient which caused the patient to go to the hospital.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive as no sample returned for evaluation.It was unknown whether the device had met relevant specifications.The product was used for urological care.It was unknown whether the product had caused the reported failure.A potential root cause for this failure mode could be due to a user related (contact with sharp objects or exposed to petrolatum based products or mechanical failure or operator error).The device was not returned for evaluation.The lot number was unknown therefore the device history record could not be reviewed.The instructions for use were found adequate and state the following: "caution: this product contains natural rubber latex which may cause allergic reactions.Sterile unless package is opened or damaged.Warning: on catheter, do not use ointments or lubricants having a petrolatum base.They will damage latex and may cause balloon burst.Store catheters at room temperature away from direct exposure to light, preferably in the original box.Single patient use only.Do not reuse and resterilize.For urological use only.Use luer slip syringe.Do not use needle.Visually inspect the product for any imperfections or surface deterioration prior to use.Do not use if package opened or damage.Recommended inflation capacities: 5cc balloon: use 10cc sterile water 30cc balloon: use 35cc sterile water do not exceed recommended capacities to deflate catheter balloon: gently insert a syringe in the catheter valve.Never use more force than is required to make the syringe stick in the valve.If you notice slow or no deflation, reseat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact and adequately trained professional for assistance, as directed by hospital protocol." the device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the foley catheter balloon burst was noted inside the patient which caused the patient to go to the hospital.
|
| |
|