• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US IRRIGATION TUBESET 24PK; SURGICAL IRRIGATION TUBING SET, GENERAL-PURPOSE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US IRRIGATION TUBESET 24PK; SURGICAL IRRIGATION TUBING SET, GENERAL-PURPOSE Back to Search Results
Model Number 284504
Medical Device Problem Code Overfill (2404)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/23/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Udi: (b)(4).The expiration date is currently unavailable.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event or Problem Description
It was reported by sales rep that during a shoulder procedure on (b)(6) 2021, it was observed that the chamber on the irrigation tubing set device filled up fully.Another pump device was used to with another tubing device.There was no surgical delay or patient consequences reported.No additional information was provided.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The device manufacture date was reported as unknown on the initial report; and has been updated accordingly.Investigation summary: the complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.With the information provided, and without the complaint device to evaluate, we cannot determine a root cause for the reported failure.A manufacturing record evaluation was performed for the finished device lot number (3000883), and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IRRIGATION TUBESET 24PK
Common Device Name
SURGICAL IRRIGATION TUBING SET, GENERAL-PURPOSE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key11629105
Report Number1221934-2021-01134
Device Sequence Number3279543
Product Code HRX
UDI-Device Identifier10886705017393
UDI-Public10886705017393
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K951843
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number284504
Device Catalogue Number284504
Device Lot Number3000883
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/24/2021
Supplement Date Received by Manufacturer05/18/2021
Initial Report FDA Received Date04/07/2021
Supplement Report FDA Received Date05/18/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-