ALCON RESEARCH, LLC - HOUSTON LAUREATE SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065750541
Medical Device Problem Code No Flow (2991)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/26/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported that there was no fluid coming out from the tubing during surgery.The product was replaced and the surgery was completed.There was no patient harm.
 
Additional Manufacturer Narrative
Additional information provided in h.3, h.6 and h.10 a review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.It was reported by the customer that ¿no fluid came through tubing".No sample has been returned for evaluation; therefore, the condition of the product could not be verified.The root cause of the customer's complaint could not be established as a sample has not been received and the condition of the product could not be verified.Without analysis of the sample, it is not possible to isolate the root cause.As the root cause is unknown, the relationship, if any, of the device to the reported incident cannot be determined.The root cause for this complaint is not known, therefore, specific action with regards to this complaint cannot be taken.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.One wet fluidics management system (fms) was visually inspected and functionally tested.No defects were observed.The ball in the drip chamber check valves moved freely per specification.The sample primed and tuned successfully.Vacuum, irrigation, and aspiration were tested at appropriate settings and met specifications.No message code, no fluid or air leaks were found on the cassette, and no defects were found.No occlusion was found in all manifolds.No leakage was detected on the irrigation and aspiration tubing insertion to the housing and fitting connection.The interface welding of the cover and the base were in good condition.The sample passed functionality testing.The root cause of the customer's complaint could not be established since the returned sample met specifications.No contributing factors could be identified that could cause the reported complaint.After the investigation of this complaint, it has been determined that this sample met specifications.Therefore, no action will be taken at this time.Quality assurance will continue to monitor customer complaints via the complaint review meetings, and will take action for any future occurrences as is deemed necessary.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAUREATE SURGICAL PROCEDURE PAK
Common Device Name
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
MDR Report Key11629431
Report Number1644019-2021-00238
Device Sequence Number14272707
Product Code LRO
UDI-Device Identifier00380657505418
UDI-Public00380657505418
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K880961
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2022
Device Catalogue Number8065750541
Device Lot Number2395543H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/21/2021
Initial Date Received by Manufacturer 04/01/2021
Supplement Date Received by Manufacturer05/28/2021
07/08/2021
Initial Report FDA Received Date04/07/2021
Supplement Report FDA Received Date05/31/2021
07/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-