| Model Number |
PS11.5-80SJ280 |
| Medical Device Problem Code |
Material Discolored (1170)
|
| Health Effect - Clinical Code |
Osteolysis (2377)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device has been returned and is pending evaluation.A supplemental report will be submitted with device evaluation findings.
|
| |
|
Event or Problem Description
|
|
Information was received that a nail was explanted due to osteolysis.It was reported the nail had reached its intended length but upon explant, the nail was found to be discolored.The patient received bone grafting and plating.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: review of the device history record found no discrepancies related to the reported event and the lot met all release criteria.A visual inspection of the returned nail revealed that the distraction rod of the nail had discoloration.The root cause is unable to be determined at this time.
|
| |
|
Event or Problem Description
|
|
No new information provided.
|
| |
|