ADVANCED BIONICS, LLC HIRES ULTRA 3D IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1601-04
Device Problems Mechanical Problem (1384); Impedance Problem (2950)
Patient Problem Failure of Implant (1924)
Event Date 05/06/2021
Event Type  malfunction  
Event Description
A review of the recipient¿s test data indicated impedance issues.Revision surgery is scheduled.
 
Manufacturer Narrative
The recipient's device was reportedly explanted.The recipient was re-implanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed cuts to the top cover and fantail revision.In addition, the electrode was severed prior to receipt.This is believed to have occurred during revision surgery.The photographic imaging inspection revealed cut electrode wires in the fantail.This is believed to have occurred during revision surgery.System lock was verified.The device passed some of the electrical tests performed.The device passed the mechanical test performed.The failure of this device is attributed to an electrode short in the electrode pocket.A corrective action was implemented.This version of the ultra device is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES ULTRA 3D IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
MDR Report Key11631592
MDR Text Key245978131
Report Number3006556115-2021-00382
Device Sequence Number1
Product Code MCM
UDI-Device Identifier07630016862250
UDI-Public(01)07630016862250(11)190225(17)220228
Combination Product (y/n)N
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup,Followup,Followup
Report Date 03/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/07/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/28/2022
Device Model NumberCI-1601-04
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/11/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age49 YR
-
-