| Model Number |
PM3210 |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/25/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
It was reported that the asymptomatic patient was dependent on the device for normal function.It was noted that an anomaly was observed whereby the elective replacement indicator (eri) was reached in 2014 but battery voltage was normal with normal device function.Remote transmission on (b)(6) 2021 had not shown eri but in person check in clinic on (b)(6) 2021 showed eri in 2014.The physician opted to change the generator since the patient was dependent.The device was explanted and replaced.The patient was stable.
|
| |
|
Search Alerts/Recalls
|