SMITH & NEPHEW, INC. GII MIS DCF AP CT BLK 3; PROSTH, KNEE, PATE/FEMOROTIBIAL, SEMI-CONST, UNCEM, PORO, COAT, POLY/METAL/POLY
|
Back to Search Results |
|
| Model Number |
71441151 |
| Medical Device Problem Codes |
Failure to Align (2522); Naturally Worn (2988)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/21/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The devices are heavily worn with many nicks, scratches, and burrs around the cutting slots.The device was manufactured in 2011 and shows signs of extensive use.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Event or Problem Description
|
|
It was reported that during a total knee arthroplasty the genesis ii mis dcf ap ct blk from size 3 to 8 are having issues fitting.No delay or injury reported.The procedure was finished using the same device.
|
| |
|
Additional Manufacturer Narrative
|
|
The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The devices are heavily worn with many nicks, scratches, and burrs around the cutting slots.The device was manufactured in 2011 and shows signs of extensive use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|