• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER DENTAL IMP,TSV,4.1,16,MTX,MG; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER DENTAL IMP,TSV,4.1,16,MTX,MG; DENTAL IMPLANT Back to Search Results
Model Number TSVT4B16
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/10/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).Gender unknown / not provided.A summary investigation has been completed for lack of primary stability events that do not allege a deficiency with the implant and identified that the reported event is likely due to biological factors which have an adverse effect on implant stability or is related to surgical technique and insertion torque.Due to a wide range of external factors (non-design/ non-manufacturing related), identifying a definitive root cause is generally not possible.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.Summary reporting.
 
Event or Problem Description
Dr.Indicated lack of primary stability, new implant will be placed within 1 month.Tooth location # 25.
 
Additional Manufacturer Narrative
Similar complaints for the inability to achieve primary stability (unable to place implant into osteotomy) have been previously investigated.Refer to attached summary investigation.Visual and dimensional evaluations of the previously returned product have not identified or suggested manufacturing non-conformances.While non-conformances were identified for some lots during manufacturing records reviews, the documented disposition actions for each have not suggested the likely release of non-conforming product.To date, all complaint data was found to be conforming and did not meet capa/hhe/d/ie escalation.Therefore, there were no complaints which confirmed a manufacturing or design related issue caused or contributed to the reported event.Dhr review was completed for the subject lot number (63838190).It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Complaint history review: complaint history review was performed for the reported lot number (63838190) for similar events and no other complaint was identified.As documented in the summary investigation, contributing factors for the reported event likely exist outside of zimmer biomet control, including those related to patient biological factors/condition and surgical technique.The following sections have been updated: b4: date of this report.D9: device availability.G3: date received by manufacturer.G6: checked "follow-up".H2: checked follow-up type.H3: changed "no" to "yes".H6: entered evaluation codes.H10: added manufacturer narrative.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
Similar complaints for the inability to achieve primary stability (unable to place implant into osteotomy) have been previously investigated.Refer to attached summary investigation.Visual and dimensional evaluations of the previously returned product have not identified or suggested manufacturing non-conformances.While non-conformances were identified for some lots during manufacturing records reviews, the documented disposition actions for each have not suggested the likely release of non-conforming product.To date, all complaint data was found to be conforming and did not meet capa/hhe/d/ie escalation.Therefore, there were no complaints which confirmed a manufacturing or design related issue caused or contributed to the reported event.Dhr review was completed for the subject lot number (63838190).It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Complaint history review was performed for the reported lot number (63838190) for similar events and no other complaint was identified.As documented in the summary investigation, contributing factors for the reported event likely exist outside of zimmer biomet control, including those related to patient biological factors/condition and surgical technique.The following sections have been updated: b4: date of this report g3: date received by manufacturer g6: checked "follow-up" h2: checked follow-up type h3: changed "yes" to "no" h6: entered evaluation codes h10: added manufacturer narrative h3 other text : device not returned.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMP,TSV,4.1,16,MTX,MG
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
MDR Report Key11633216
Report Number0002023141-2021-00889
Device Sequence Number3356281
Product Code DZE
UDI-Device Identifier00889024019850
UDI-Public(01)00889024019850(17)221130(10)63838190
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K101977
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Dentist
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/30/2022
Device Model NumberTSVT4B16
Device Catalogue NumberTSVT4B16
Device Lot Number63838190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/06/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/15/2021
Supplement Date Received by Manufacturer06/30/2021
08/13/2021
Initial Report FDA Received Date04/08/2021
Supplement Report FDA Received Date07/02/2021
08/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
Patient Weight85
-
-