BIOTRONIK AG, BUELACH, SWITZERLAND ORSIRO 3.5/35; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number 391018
Medical Device Problem Code Deflation Problem (1149)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/16/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Combination product: yes.
 
Event or Problem Description
An orsiro drug-eluting stent system was selected for treatment of a pre-dilated lesion (75 percent stenosis degree) in the rca.After stent implantation it was noticed that the deflation time was longer than usual (more than 15 seconds, normally 10 seconds).
 
Additional Manufacturer Narrative
Combination product: yes the returned instrument was subjected to a technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The inflation device used in the intervention was returned and identified as a medtronic everest 30.The technical investigation revealed that the balloon has been inflated and was deflated in the as-returned state.A small amount of contrast medium residue was observed in the balloon- and inflation lumen.The balloon could be inflated and deflated within the specified maximum deflation time with the returned inflation device medtronic everest 30.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.Based on the conducted investigations, no manufacturing or material related root cause could be determined.The root cause for the reported event is most likely related to the handling during the procedure (i.E.Insufficient deflation time).It should be noted that the ifu advises the user that the instrument shall be deflated for at least 35 seconds.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORSIRO 3.5/35
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key11633475
Report Number1028232-2021-01661
Device Sequence Number14413872
Product Code NIQ
UDI-Device Identifier07640130439399
UDI-Public07640130439399
Combination Product (Y/N)Y
Initial Reporter CountryCH
PMA/510(K) Number
P170030
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/17/2022
Device Model Number391018
Device Catalogue NumberSEE MODEL NO.
Device Lot Number03204787
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/06/2021
Initial Date Received by Manufacturer 04/07/2021
Supplement Date Received by Manufacturer06/01/2021
Initial Report FDA Received Date04/08/2021
Supplement Report FDA Received Date06/01/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-