| Model Number |
OM-1502 |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/1901
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during procedures one doctor has been having an issue with the magnum 2.8 anchors.He has had a string of implants that haven't been releasing well after implantation.In some cases, he has needed to really pull on the inserter to get to detach or even use a mallet to back it out.The exact amount of times this has happened, number of devices, patients and procedures involved is unknown.Unknown if there was any delay or backup available.No patient injuries or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported device used in treatment, was not returned to the designated complaint unit for independent evaluation.Thus visual inspection and functional testing could not be performed.A review of device records showed, there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded, this was an isolated event.The instructions for use was reviewed and found to include conditions of off label use.And technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files, found that the reported failure was documented appropriately.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|
| |
|
Search Alerts/Recalls
|