| Model Number |
DAQ916 |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/02/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
A health care provider reported that the nerve monitoring pre amplifier was not working properly.There was no patient impact.
|
| |
|
Event or Problem Description
|
|
On follow up it was reported that there was no data regarding stimulation and was discovered prior using on a patient.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: analysis found the customers complaint has been confirmed the device does not work properly.Channels 14, 27 and 28 are faulty.The connector seals are torn.The top decal was worn.The analog pcba has been replaced this solved the faulty channels issue.The connector seals and top decal were replaced.Mandatory parts are also replaced.The device has been successfully tested and passed according to manufacturing specifications.H6: fdm b17; fdc d14; fdr c20 are no longer applicable medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|