| Model Number |
71577-75 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/24/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product has been returned, and the investigation is currently in process.A supplemental report will be submitted once all investigation activities are complete.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Event or Problem Description
|
|
A customer reported receiving erroneous glucose results from an abbott diabetes care device.The results when plotted on a parkes error grid fell into either the c, d, or e zone.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|
Event or Problem Description
|
|
A customer reported receiving erroneous glucose results from an abbott diabetes care device.The results when plotted on a parkes error grid fell into either the c, d, or e zone.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Reader mcga217-j4248 has been returned and investigated with retained test strips.Visual inspection has been performed on the reader and no issues were observed.Control solution testing was performed and no issues were performed.All results were within range specification.No malfunction or product deficiency was identified.Extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Dhr for the libre reader indicated by the serial number provided by the customer.The dhr showed the libre reader passed all tests prior to release.Dhr for the precision strips indicated by the serial number provided by the customer.The dhr showed the precision strips passed all tests prior to release.Retain testing was performed for precision strips and all units performed in specification and passed.If the partial product is returned, a physical investigation will be performed, and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|