Model Number SN6AT6
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Codes Visual Impairment (2138); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Evaluation summary: the product was not returned for analysis.The product history and batch records were reviewed and documentation indicated the product met release criteria.The product investigation could not identify a root cause.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A facility representative reported that after the initial intraocular lens (iol) implant procedure, a patient complained of "rotated off axis, myopic result" and a clinical reason of "rotated off axis, myopic result".This required the removal of the iol in a secondary surgery where the lens was replaced with a company toric lens.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ TORIC SINGLEPIECE IOL
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
MDR Report Key11634006
Report Number1119421-2021-00772
Device Sequence Number14366969
Product Code MJP
UDI-Device Identifier00380652274838
UDI-Public00380652274838
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/08/2025
Device Model NumberSN6AT6
Device Catalogue NumberSN6AT6U185
Device Lot Number15075483
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/19/2021
Initial Report FDA Received Date04/08/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-