| Model Number |
PED-425-30 |
| Medical Device Problem Codes |
Break (1069); Material Deformation (2976); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Codes |
Low Blood Pressure/ Hypotension (1914); Stenosis (2263)
|
| Date of Event |
03/16/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received a report that the pipeline experienced resistance in the proximal section of the marksman catheter.The devices were replaced, and the patient did not experience any injury or complications.The devices were prepared and flushed according to the instructions for use (ifu).The patient was undergoing treatment for an unruptured, amorphous aneurysm.The max diameter was 6.34mm, and the neck diameter was 5.18mm.The patient's vessel tortuosity was severe.The landing zone was 4.56mm distal and 3.11mm proximal.The access vessel diameter was 3.12mm. dual antiplatelet treatment was administered.Additional information received 2021-03-22 indicated that the patient's blood vessel was extremely tortuous/twisted, and the catheter had accordioned after trying to pass through key parts many times.This caused the pipeline to fail to be delivered during the deployment process.At that time, the patient's blood vessel dropped, and the operation could not be performed.Due to vascular stenosis, only one stryker intracranial stent was used to treat the stenoisis, and the aneurysm operation was not successful.The pipeline pushwire was kinked/separated, but images showed no separation of pushwire.
|
| |
|
Event or Problem Description
|
|
Additional information received reported there was no medical intervention required to treat the patient's drop in blood pressure.T here was no additional surgery scheduled or planned to complete treatment of the patient's aneurysm.It was clarified that the pipeline pushwire did not actually separate or break.The patient was recovering from the procedure with assistance of medication.
|
| |
|
Additional Manufacturer Narrative
|
|
A4.Patient weight (74kg) updated b5.Updated with additional information received.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: analysis of the pipeline embolization device (lot no.B075766) found that the pushwire was extending ~48.6 cm from hub and ~4.9 cm from distal tip.An attempt was made to push the pipeline flex device further from the marksman catheter; however, due to high resistance the pipeline flex was unable to be pushed out.No bends or kinks were found with the pipeline pushwire.The distal hypotube and ptfe were found to be intact.The proximal bumper and re-sheathing pad were found to be intact.The pad restraint was found to be damaged.The dps sleeves were found to be intact and damaged.The tip coil was found to be intact.The proximal braid end was found to be opened.The distal braid end was found to be opened and frayed.No other damages or anomalies were found with the device.Based on the device analysis and reported information, the customer¿s report of ¿resistance/stuck during delivery¿ was confirmed.Based on the investigation conducted resistance can occur during tracking, deployment and re-sheathing of the device in distal and tortuous anatomies.In addition, resistance can occur due to patient vessel tortuosity, failure to maintain a continuous flush, or pipeline is pulled back/torqued during delivery.The vessel anatomy was moderate in tortuosity and a continuous flush was maintained.Therefore, the cause could not be determined.H6: method code updated to b01.Result code updated to c0702 and c070601.Conclusion code updated to d15.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|