Catalog Number 096002-000-591 |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/25/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
|
|
Event Description
|
The account alleges that during the hemodynamic monitoring of a patient, the pressure monitoring [pm] set was found to be leaking fluid at the blood sampling septum.No patient injury to report.
|
|
Manufacturer Narrative
|
The suspect device has been returned for evaluation.The complaint is confirmed.The root cause is attributed to the manufacturing process.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.
|
|