| Model Number |
5640 |
| Medical Device Problem Code |
Separation Problem (4043)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
03/12/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Manufacturing records will be reviewed.A follow-up report will be issued after the investigation is complete.
|
| |
|
Event or Problem Description
|
|
As reported: physician used a turnpike spiral and during the procedure a piece of the black tip separated from the end and was in the left main.They were unable to retrieve using a snare, so they placed a stent in the left main to keep the piece of the tip in place against the wall of the artery.
|
| |
|
Additional Manufacturer Narrative
|
|
Turnpike was returned to vsi/teleflex for evaluation.Lot: 704744 was reviewed.There are no non-conformances related to this lot therefore supporting the device met material, assembly and performance specifications.Additional information was requested form the account.No response was received.Only an image of the turnpike was sent confirming damages to the tip.The tip was unable to be snared out of the vessel.A stent was placed in the left main to keep the piece of the tip in place against the wall of the artery.No blood particulates were noted on the shaft.The tip of the catheter was severely damaged, torqued, and fatigued.Approximately 0.2 cm of tip material was observed to be sheared off from the distal section of the tip.The shaft of the catheter had a minor kink.Length, o.D., o.D at the distal section, were measured and catheter met specification per drawing.No other damages were noted.
|
| |
|