| Model Number |
TRL-350 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a left side rupture.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: d9, h3, h6.Device evaluation: visual analysis of the returned device identified: opening, weight to the spec, non-penetrating nick, crease flat, black particles on the surface of the shell.A microscopic analysis was performed which identify sharp edge and with non-penetrating nicks assessed as surgical impact opening.A dimension measurement in the shell was performed which identify the thickness within specification.Based on the device analysis the final assessment is: a sharp edge opening with non-penetrating nicks assessed as surgical impact.Non-penetrating nicks.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a left side rupture.Device has been explanted.
|
| |
|