• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD Back to Search Results
Model Number 400SMTHXSFT0204
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/09/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Evaluation of the first returned smart coil revealed that the pusher assembly mid-joint was retracted through the introducer sheath friction lock.If this occurs, resistance may be experienced during advancement.Further evaluation revealed that the friction lock was wrinkled, and the pusher assembly had bends and kinks along its length.The wrinkled friction lock was likely a result of forceful gripping of the friction lock during manipulation against resistance.The pusher assembly kinks were likely incidental to the reported complaint.During functional testing, the pusher assembly mid-joint was advanced through the introducer sheath friction lock with resistance.The smart coil was then advanced through the introducer sheath with resistance due to pusher assembly kinks.Evaluation of the second returned smart coil revealed that the pusher assembly mid-joint was retracted through the introducer sheath friction lock.If this occurs, resistance may be experienced during advancement.During functional testing, the pusher assembly mid-joint was advanced through the introducer sheath friction lock with resistance.The smart coil was then advanced through the introducer sheath without an issue.Penumbra coils are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2021-00713.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the basilar tip using penumbra smart coils (smart coils) and a non-penumbra microcatheter.During the procedure, the physician placed three coils into the aneurysm using the microcatheter.Next, the hospital technologist attempted to advance a smart coil out of its introducer sheath; however, the smart coil would not advance at all within its introducer sheath.Subsequently, the physician flushed the introducer sheath and attempted to re-advance the smart coil within its introducer sheath but was unsuccessful.Therefore, the smart coil was removed.The physician then successfully placed a new smart coil into the aneurysm using the microcatheter.Afterwards, the physician attempted to advance another smart coil within its introducer sheath; however, the same issue occurred.The smart coil would not advance within its introducer sheath.It was reported that the introducer sheath of the smart coil was stuck on its pusher assembly.Therefore, the smart coil was removed.The procedure was completed using two additional smart coils.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key11638615
Report Number3005168196-2021-00714
Device Sequence Number3546623
Product Code HCG
UDI-Device Identifier00814548015712
UDI-Public00814548015712
Combination Product (Y/N)Y
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K160832
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/11/2021
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/27/2025
Device Model Number400SMTHXSFT0204
Device Catalogue Number400SMTHXSFT0204
Device Lot NumberF97573
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/22/2021
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/11/2021
Initial Report FDA Received Date04/08/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age74 YR
Date Report Sent to FDA01/01/2005
-
-